Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 65
Test 7a: (Q) Pulse Burst Frequency - 53230A Channel 3
Option 150 With Option 106
This test uses the 53230A to measure both narrow and wide pulse outputs
of a 6 GHz carrier.
NOTE This test procedure applies only to the 53230A Counter with
Option 106 and Option 150 installed. Option 106 is installed as
microwave channel 3.
NOTE The Quick Verification (Q) requires only the Narrow Pulse Width
Burst Frequency Test be performed.
Equipment
FS725 Rubidium Timebase
Agilent E8257D-520-UNW Microwave Signal Generator
Agilent 10503A BNC Cable (2)
Agilent 11500E (3.5) mm cable
Agilent 1250-1250 N(M) to SMA(F)
E8257D-520-UNW Setup
Output Frequency: 6 GHz
•Output Power: 0 dBm
Auto-Level Control: Off
•Pulse Mode: On
Pulse Repetition Interval Period: 1 mSec
Pulse Width (Narrow): 200 nSec
NOTE The input power MUST be 0 ± 0.5 dBm at the channel 3 input
connector. If necessary, use a 3.5 mm cable with certified loss at 6
GHz or a power meter to verify input power.