Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
58 Assembly-Level Service Guide
6 Set the E8257D-520 to 5.88 GHz at -27 dBm.
NOTE The input power MUST be -27 ± 0.5 dBm at the channel 3 input
connector. If necessary, use a 3.5 mm cable with certified loss at
5.88 GHz or a power meter to verify input power.
7Press the Trigger key.
The counter takes 100 readings and fills in the math functions.
8 Verify the ‘Mean’ is:
5.88 GHz ± 58.800 Hz (53210A)
5.88 GHz ± 5.880 Hz (53220A/53230A)
9 Verify the ‘Std Dev’ is less than:
82.300 Hz (53210A)
41.100 Hz (53220A)
8.230 Hz (53230A)
10 Record the Mean and Standard Deviation for 5.88 GHz, Test 6a, in
the Agilent 53210A/53220A/53230A Performance Test Record.
11 Press the Reset Stats softkey.
12 If Opt 150 is installed, proceed to Test 7a. Otherwise, disconnect
the test setup.