Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 57
11 Press to set:
Statistics: On
The display shows:
Channel 3 (Option 106) Frequency Accuracy Test Procedure
1Press the Trigger key.
The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
100 MHz ± 1.000 Hz (53210A)
100 MHz ± 100 mHz (53220A/53230A)
3 Verify the ‘Std Dev’ is less than:
1.400 Hz (53210A)
700 mHz (53220A)
140 mHz (53230A)
4 Record the Mean and Standard Deviation for 100 MHz, Test 6a, in
the Agilent 53210A/53220A/53230A Performance Test Record.
5Press the Reset Stats softkey.