Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 55
NOTE For the Agilent 53210A Counter, Option 106 is installed as
Channel 2. For the Agilent 53220A and 53230A Counters, Option
106 is installed as Channel 3. For this procedure, “Channel 3” is
used to refer to the microwave input channel.
3 Press the Channel 3 input key.
Channel 3 is selected for test input.
NOTE There are no parameters to set for Channel 3 for this test.
4Press the Gate key.
•The Gate softkeys are displayed:
5 Press and set:
Gate Src: Timed
Gate Out: On
•Gate Time: 10 mSec