Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
54 Assembly-Level Service Guide
E8257D-520 Setup
Output Frequency: 100 MHz
•Output Power: -27 dBm
NOTE The input power MUST be -27 ± 0.5 dBm at the channel 3 input
connector. If necessary, use a 3.5 mm cable with certified loss at
100 MHz or a power meter to verify input power.
Counter Setup
1Press the Preset key to preset the counter.
2 Connect the equipment as shown in Figure 1-8.
Figure 1-8. Option 106 ( 6 GHz ) Microwave Channel 2/3 Test Setup
Opt 010 UO CXO
Lin e
100-240V, 50-60 Hz
100-127V, 400 Hz
90VA M ax
U S B L AN
In t R e f O ut T rig In
E xt R ef In G ate In /O ut
Ch 1 Opt 20 1 C h 2 G P-IB
Opt 106/115
IECS / NMB-001
N10149
ISM 1-A
C U S
Opt 106
100 MH z - 6 GHz
+27 dBm Max
!
Opt 300 B attery
Opt 115
300 MH z - 15 GHz
+27 dBm Max
!
FS725
Rubidium
Timebase
Ext
Ref
In
E8257D-520
Microwave
Signal
Generator
Ch 2/3