Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 53
Time Interval Test Procedure
1Press the Trigger key.
The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
49.9 ± 0.5 nSec (53220A)
49.9 ± 0.25 nSec (53230A)
3 Verify the ‘Std Dev’ is less than:
.140 nSec (53220A)
.028 nSec (53230A)
4 Record the Median and Standard Deviation, Test 5, in the “Agilent
53210A/53220A/53230A Performance Test Record (Tests 1 to 7)
starting on page 78.
5 Proceed to Test #6 or disconnect the test setup.
Test 6a: (Q) Frequency Accuracy - Channel 2/3, Option
106 (6 GHz Microwave Channel)
This test verifies the frequency accuracy of Option 106 Channel 2
(53210A) or Channel 3 (53220A/53230A) at two frequencies (100 MHz and
5.88 GHz).
Equipment
FS725 Rubidium Timebase
Agilent E8257D-520 Microwave Signal Generator
Agilent 10503A BNC Cable (2)
Agilent 11500E (3.5) mm cable
Agilent 1250-1250 N(M) to SMA(F)