Agilent 53210A/53220A/53230A Complete Performance Tests
50 Assembly-Level Service Guide
2Press the Preset key to preset the counter.
Display shows:
Instrument has been preset.
Ext Ref is shown in the upper right-hand corner of the display.
3Press the Time Interval key.
The following softkeys are displayed:
4Press the Time Interval softkey.
The Time Interval softkeys are displayed:
5 Press to set Time Int: 1-2.
NOTE Time Int 2-1 could also be tested, but only one test is required. If
Time Int 2-1 is used, set the 81134A to delay Channel 1 by 49.9 nSec
instead of Channel 2.
6Press the Start Ch Level softkey.
7 The following softkeys are displayed: