Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 49
CAUTION The 3.5 mm cables must be the exact same length or erroneous
measurements will occur.
81134A Setup
•Output Level: 1 Vpp
Period: 66.6 nSec
Pulse Width Ch1: 5 nSec
Pulse Width Ch2: 5 nSec
•Delay Ch2: 49.9 nSec
Counter Setup
1 Connect the equipment as shown in Figure 1-7.
Figure 1-7. Time Interval (Channel 1 to 2) Test Setup
Opt 010 UO CXO
Lin e
100-240V, 50-60 Hz
100-127V, 400 Hz
90VA M ax
U S B L AN
In t R e f O ut T rig In
E xt R ef In G ate In /O ut
Ch 1 Opt 20 1 C h 2 G P-IB
Opt 106/115
IECS / NMB-001
N10149
ISM 1-A
C U S
Opt 106
100 MH z - 6 GHz
+27 dBm Max
!
Opt 300 B attery
Opt 115
300 MH z - 15 GHz
+27 dBm Max
!
FS725
Rubidium
Timebase
Ext
Ref
In
81134A
Pulse
Generator
Ch 1 Ch 2