Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
48 Assembly-Level Service Guide
0.140 nSec (53220A)
0.028 nSec (53230A)
4 Record the Mean and Standard Deviation for Channel 1, Test 4a,
in the “Agilent 53210A/53220A/53230A Performance Test Record
(Tests 1 to 7)” starting on page 78.
5 Repeat steps 1 to 3 for Channel 2 (Test 4b). Be sure to do the
Channel 2 setup first. (See steps 8, 9 in the Counter setup above.)
NOTE Selecting Channel 2 automatically resets the statistics.
6 Record the Mean and Standard Deviation for Channel 2, Test 4b,
in the “Agilent 53210A/53220A/53230A Performance Test Record
(Tests 1 to 7)” starting on page 78.
7 Proceed to Test #5 or disconnect the test setup.
Test 5: (Q) Time Interval
This test verifies the Time Interval specifications between Channel 1 and
Channel 2.
NOTE Test #5 applies to the 53220A and 53230A only.
Equipment
FS725 Rubidium Timebase
Agilent 81134A Pulse Generator
Agilent 10503 BNC Cables (2)
Agilent 11500E (3.5) mm cables (2)
Agilent 1250-1200 BNC(f)-to-SMA(m) Adapter