Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 43
13 Record the ‘Mean’ and ‘Std Dev’ for 1 MHz and 250 MHz, Channel
2, Test 3c and 3d, in the Agilent 53220A/53230A Performance Test
Record starting on page 78.
14 Press the Reset Stats softkey.
15 Proceed to Test #4 or disconnect the setup.
Test 4: (Q) Pulse Width
This test uses the Agilent 53220A/53230A Counter to measure the width
of a narrow pulse from a pulse generator.
NOTE Test #4 applies to the 53220A and 53230A only.
Equipment
FS725 Rubidium Timebase
Agilent 81134A Pulse Generator
Agilent 10503 BNC Cables (3)
Agilent 15443A (SMA) cable or Agilent 11500E (3.5 mm) cable
Agilent 1250-1200 BNC(f)-to-SMA(m) Adapter (2)
81134A Setup
•Output Level: 1 Vpp
Period: 66.6 nSec
Pulse Width: 5 nSec