Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
38 Assembly-Level Service Guide
Equipment
FS725 Rubidium Timebase
CG635 Synthesized Signal Generator (E8257D-520 is a valid substitute)
Agilent 10503 BNC Cables (3)
CG635 Signal Generator Setup
1 Set Output Channel: Q
2 Set Output Level: +7 dBm
3 Set Frequency: 1 MHz
Counter Setup
1 Connect the equipment as shown in Figure 1-5.
Figure 1-5. Test Setup for Single Period
2Press the Preset key to preset the counter.
Opt 010 UO CXO
Lin e
100-240V, 50-60 Hz
100-127V, 400 Hz
90VA M ax
US B L AN
Int Re f Ou t Trig In
Ex t Re f In Ga te In /Ou t
Ch 1 Opt 20 1 Ch 2 GP -IB
Opt 106/115
IECS / NMB-001
N10149
ISM 1-A
C U S
Opt 106
100 MHz - 6 GHz
+27 dBm M ax
!
Opt 300 B attery
Opt 115
300 MHz - 15 GHz
+27 dBm Max
!
FS725
Rubidium
Timebase
Ext
Ref
In
CG635
Signal
Generator
Ch 1