Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
36 Assembly-Level Service Guide
3 Verify the ‘Std Dev’ is less than:
140 mHz (for 53210A)
70 mHz (for 53220A)
14 mHz (for 53230A)
4 Record the ‘Mean’ and ‘Std Dev’, Test 2a, for Channel 1 (10 MHz),
in the “Agilent 53210A/53220A/53230A Performance Test Record
(Tests 1 to 7)” starting on page 78.
5Press the Reset Stats softkey.
6 Set CG635 output frequency to 115 MHz and repeat step 1.
7 Verify the ‘Mean’ is:
115 MHz ± 1.150 Hz (for 53210A)
115 MHz ± 0.115 Hz (for 53220A/53230A)
8 Verify the ‘Std Dev’ is less than:
1610 mHz (for 53210A)
805 mHz (for 53220A)
161 mHz (for 53230A)
9 Record the ‘Mean’ and ‘Std Dev’, Test 2a, for Channel 1 (115 MHz),
in the “Agilent 53210A/53220A/53230A Performance Test Record
(Tests 1 to 7)” starting on page 78.
10 Press the Reset Stats softkey.
11 Set CG635 output frequency to 350 MHz, and repeat step 1.
12 Verify the ‘Mean’ is:
350 MHz ± 3.5 Hz (for 53210A)
350 MHz ± 0.350 Hz (for 53220A/53230A)