Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 35
15 Press the Statistics softkey.
–The Statistics softkeys are displayed:
16 Press to set:
Statistics: On
The display shows:
Frequency Accuracy Test Procedure
1 Press the Trigger key to manually trigger the counter.
Counter takes 100 readings and fills in Math functions.
2 Verify the ‘Mean’ is:
10 MHz ± 100 mHz (for 53210A)
10 MHz ± 10 mHz (for 53220A/53230A)