Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 33
The following softkeys are displayed:
6 Press and set:
Noise Rejection: Off
7Press the Trigger key.
The following softkeys are displayed:
8 Press and set:
Trigger Source: Manual
Wait For Trig is displayed.
•Rdgs/Trigger: 100.
9Press the Gate key.
•The Gate softkeys are displayed:
10 Press and set:
Gate Src: Timed
•Gate Time: 10 mSec
Gate Out: On
Polarity: Norm