Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
30 Assembly-Level Service Guide
Absolute Timebase Test Procedure
1Press the Trigger key to manually trigger the counter.
Counter takes 100 readings and fills in Math functions.
2 Verify the ‘Mean’ displays a reading of:
10 MHz ± 15 Hz (for Standard Timebase)
10 MHz ± 0.505 Hz (for Option 010 High Stability Timebase)
3 Record the ‘Mean’ test value, Test 1, in the “Agilent
53210A/53220A/53230A Performance Test Record (Tests 1 to 7)
starting on page 78.
4 Proceed to Test #2 or disconnect the test setup.
Test 2: (Q) Frequency Accuracy - Channels 1 and 2
This set of tests verifies the Channel 1 and 2 frequency accuracy
specifications of the Agilent 53210A/53220A/53230A Frequency Counters.
Equipment
FS725 Rubidium Timebase
CG635 Synthesized Signal Generator (E8257D-520 is a valid substitute)
Agilent 10503A BNC Cables (3)
CG635 Signal Generator Setup
1 Set Output Channel: Q
2 Set Output Level: +7 dBm
3 Set Frequency: 10 MHz