Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 27
2Press the Preset key on the front panel to preset the counter.
Display shows:
Instrument has been preset.
Channel 1 annunciator is lit.
3Press the Trigger key.
The following softkeys are displayed:
4 Press and set:
Trigger Source: Manual
Wait For Trig is displayed.
•Rdgs/Trigger: 100.
5Press the Freq Period key.
The following softkeys are displayed:
6 Press the Freq softkey.
The following softkeys are displayed:
*
*Opt 150