Specifications

Table Of Contents
Chapter 6 Backdating
Procedure Changes
Assembly-Level Service Guide 257
NOTE If you do not have a remote computer available (requires a LAN,
USB or GPIB interface) or are not proficient at programming,
contact the nearest Agilent Service Center for information and
help on sending the counter to Agilent for repair and/or
calibration. Sse the website information on page 2.
19 If you are able to read the error message(s), consult “Error
Messages” starting on page 163 for what to do after receiving a
specific calibration error message.
Summary of Option 106 Microwave Calibration Procedure:
Input signal: This should be a clean sine wave at each frequency and
power level, driving 50 termination.
Timebase: Any external timebase you provide is ignored during
calibration.
Verify signal input connections and source output is ON.
Procedure: From the front-panel, the following key strokes initiate
the microwave frequency channel calibration:
[Utility] / (Cal AutoCal) / (Perform Cal) / (Cal Step uWave) / enter (Input
Freq) <frequency>
[Utility] / (Cal AutoCal) / (Perform Cal) / (Cal Step uWave) / enter (Input
Power) <power> / (Perform Cal Step)
Summary of Option 106 Remote Microwave Calibration :
Set timeout for PC to be greater than 20 seconds.
Connect calibration source to channel 2 (53210A) or channel 3
(53220A/53230A) at 500 MHz, -27 dBm for Option 106).
Unlock security: CALibration:SECurity:STATe OFF, <code>
Read security state: CALibration:SECurity:STATe?
0 means OFF or unlocked.
For the 53220A or 53230A Counters, send the command to calibrate
as: “CALibration:LEVel3? 500E8, -27”.