Assembly-Level Service Guide 255
Display softkey annunciators show:
11 Press the Cal Step softkey.
Display softkey annunciators show:
* Only displayed if Option 106 or 115 is installed.
12 Press the uWave softkey.
Display softkey annunciators show:
13 Use the front panel knob and left/right arrow keys to enter the
value as: 500 MHz in the data entry area to agree with your
source.
NOTE Alternatively, you may use the Shift key on the front panel to put
the keyboard keys into a numeric mode, and then enter the entire
frequency manually. The data entry area will repeat whatever
numbers are used to select the calibration frequency.
14 Press the Input Power softkey and enter the power level of the 500
MHz source as -27 dBm. (Must be -27 dBm ± 0.5 dB at the input
connector. If necessary, verify with a power meter.)
**