Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 25
Agilent 53210A/53220A/53230A Complete
Performance Tests
The following seven performance tests, when successful, verify that the
counter meets the instrument specifications as provided on the CD ROM
that came with your Agilent 53210A/53220A/53230A Counter. See also
the information on running a ‘QUICK’ performance test on page page 13.
Before running these tests, review the “Test Considerations” on page 14..
Record the results of the performance tests in the appropriate place on the
“Agilent 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)”
starting on page 78.
To verify the total number of accuracy places for frequency
measurements, it may be necessary to turn off Auto-Digit control. Use this
set of key depressions to turn AutoDigits OFF:
[Digits] / (AutoDigits) / <Off>
NOTE To perform valid testing of the specifications, warm up the
counter and test equipment for at least 30 minutes.
Table 1-2. The Agilent 53210A/53220A/53230A Performance Tests
*
Test Description
* Other counter measurement functions are mathematically derived by the processor from the
parameters verified by these performance tests.
Test 1: Absolute Time Base
Test 2: Frequency Accuracy - Channels 1 and 2
Test 3: Single Period
Test 4: Pulse Width
Test 5: Time Interval
Test 6: Frequency Accuracy - Microwave Channel Options 106 and 115
Test 7: Pulse Burst Frequency - 53230A Channel 3 Option 150