Specifications

Table Of Contents
Chapter 4 Retrofitting Options
To Retrofit the 53220A/53230A Channel 3 Input, 6.0 GHz (Option 106) or
15.0 GHz (Option 115) plus Front Panel Option 202
230 Assembly-Level Service Guide
connector of the semi-rigid cable installed on the front panel
assembly.
25 Line up the male semi-rigid SMA cable connector and the Channel
3 assmbly female SMA edge connector, connect them together and
loosely tighten the SMA connector.
NOTE Be sure the flange on the front of the Channel 3 assembly is
positioned over the tab on the front panel. Refer to Figure 4-20.
26 Re-install the T20 TORX screw removed in step 22 above, securing
the rear flange on the Channel 3 bracket to the motherboard with
the screw. Do not tighten the screw all the way down at this time.
NOTE If necessary, turn the chassis over and, through the open area
provided, loosen the SMA connector on the semi-rigid cable where
it connects to the front panel and carefully re-position the cable
until it connects the assembly to the front panel, with no stress on
the semi-rigid cable.
27 Use a 5/16" open-end wrench to secure the two SMA connectors at
each end of the semi-rigid cable. Do not overtighten.
28 Tighten the T20 TORX screw (step 26 above).
29 Install the two supplied flathead screws through the holes in the
left-hand aluminum side panel into the embedded nuts on the
bracket of the Channel 3 assembly.
30 If removed in Preliminary Step 5, re-install the Option 300
Internal Battery Assembly. Be sure to re-connect the cables.
31 Re-install the cover by performing the cover removal procedure,
in the Preliminary Steps above, in reverse.
32 Install the supplied Option 106 or 115 label alongside the front
panel TypeN connector in the area provided.