Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Operational Verification
22 Assembly-Level Service Guide
NOTE If the counter has Option 201 (rear panel input connectors)
installed, connect the Int Ref Out to the rear panel Channel 1 input
connector and NOT the front panel Channel 1 input connector.
2Press the Preset key to preset the counter.
3 Verify the display shows a frequency reading of approximately
10,000,000 Hz.
4 Mark Pass or Fail in the “Agilent 53210A/53220A/53230A
Performance Test Record (Tests 1 to 7)” starting on page 78.
NOTE The remaining steps apply only to the Agilent 53220A and Agilent
53230A since the Agilent 53210A does not have a second RF
frequency input.
5 In the same way as for Channel 1, connect the Int Ref Out to the
Channel 2 input connector (front or rear panel).
6 Press the Channel 2 button.
NOTE If necessary, press the Freq/Period key to read the Channel 2
frequency.
Channel 1 annunciator is not lit. Channel 2 annunciator is lit.
7 Verify the display shows a frequency reading of approximately
10,000,000 Hz.
8 Mark Pass or Fail in the “Agilent 53210A/53220A/53230A
Performance Test Record (Tests 1 to 7)” starting on page 78.
9 Disconnect the test setup.