Specifications

Table Of Contents
Chapter 4 Retrofitting Options
To Retrofit the 53210A (Channel 1) or 53220A/53230A (Channels 1/2)
Rear Panel Inputs (Option 201)
Assembly-Level Service Guide 205
6 Insert the BNC connector end of the RF cable (over the top of the
Option 106/115 assembly) through the appropriate channel cutout
hole in the rear panel.
7 Using a 14 mm deep-socket or spintite, secure the connector with
the included hex nut and lockwasher (do not overtighten).
8 If necessary, repeat steps 2 through 7 for Channel 2.
9 Install the supplied cable clamp(s) as shown in Figure 4-7.
NOTE When Option 201 is installed (Channels 1 and 2 rear panel BNC
connectors), the front panel connector performance for RF
Channels 1 and 2 is not specified. The rear panel connector for
each channel IS specified according to the information in Chapter
7 "Specifications".
10 If the Option 300 Internal Battery was removed at the beginning
of this procedure, re-install it now. Be sure to re-connect the
cables.
11 Re-install the cover by performing the cover removal procedure
(Preliminary Steps 1-4), in reverse.