Specifications

Table Of Contents
Chapter 4 Retrofitting Options
To Retrofit the GPIB Connector
194 Assembly-Level Service Guide
To Retrofit the GPIB Connector
GPIB Connector Parts (P/N 53200U-400)
Item Agilent Part No. Qty.
PC Board Assembly P0001-63001 1
Ribbon Cable, 16-pin 53200-61601 1
7mm Hex-head standoff 0380-0644 2
Washer, lock 2190-0577 2
Preliminary Steps
CAUTION Before proceeding with the installation, follow all the ESD steps
as outlined in the section titled "Do This First" under the
"Introduction" at the beginning of Chapter 4 of this service guide.
1 Remove the cover as follows:
NOTE Refer as necessary to the figures in the section titled “To Remove
the Cover and Rear Bezel” on page 173.
2 To remove the counter handle, rotate the handle to the vertical
(90 degree) position and pull outward on the two side arms.
3 Using a T20 TORX screwdriver, loosen the two T20 TORX screws
securing the rear panel bezel.
4 Slide the outside cover of the counter to the rear and remove the
cover.
5 If installed, disconnect and move the Option 300 Internal Battery
DC Power Assembly out of the way as follows (refer to the figures
in “To Remove the Internal DC Battery Assembly (Option 300)” on
page 175, as needed):