Specifications

Table Of Contents
Chapter 3 Replacing Assemblies
To Remove the Internal DC Battery Assembly (Option 300)
Assembly-Level Service Guide 175
To Remove the Internal DC Battery
Assembly (Option 300)
1 Remove the cover and rear bezel from the counter as described
previously.
2 Remove the two T20 TORX screws from the left hand side of the
counter that secure the battery to the frame.
3 Lift upwards on the two tabs holding the battery assembly on top
of the left side panel and slide the battery assembly off the three
large tabs that hold it against the AC Power Supply Assembly.
4 Without disconnecting the battery charging cable, carefully turn
the battery assembly over on its back and insert the three smaller
tabs behind the large tabs into the three slots in the assembly.
(See Figure 3-3.)
5 Disconnect the Y-Cable (3 connectors) from the Option 300
Battery Assembly, then disconnect the digital cable.
6 Remove the Battery Assembly. If the battery is to be replaced,
remove the four T10 TORX screws holding the battery cover.
7 Lift the battery and its cover free of the battery connector. Push
the battery out of the aluminum cover. Insert the new battery in
the cover and re-install the battery into the charging assembly.
8 Re-install the new battery and/or charging assembly into the
counter by reversing the above procedure. Be sure to re-connect
the cables to the assembly, digital cable first.