Specifications

Table Of Contents
Chapter 2 Service
Error Messages
166 Assembly-Level Service Guide
Query Errors (-400...)
These errors indicate licensing issues within the counter firmware. Refer
to the User’s Guide for information.
+600, “Internal licensing error”
+601, “License file corrupt or empty””
+602, “No valid licenses found for this instrument”
+603, “Some licenses could not be installed”
+604, “License not found”
Calibration Errors (700...)
The following errors indicate failures that may occur during a
calibration and which are usually caused by the operator or by
the external calibration source.
NOTE If unable to determine the cause of a calibration error, run the
counter Self Test to determine the instrument status:
[Utility] / (Instr Setup) / (Self Test)
+701, “Calibration error; security defeated”
+702, “Calibration error; calibration memory is secured”
+703, “Calibration error; secure code provided was invalid”
+704, “Calibration error: secure code too long”
+705, “Calibration error; calibration aborted”
+706, “Calibration error: provided value out of range”
+707, “Calibration error: computed correction factor out of range”
+708, “Calibration error: signal measurement out of range”
+709, “Calibration error: no calibration for this function”
+710, “Calibration error: full scale correction out of range”
+711, “Calibration error: calibration string too long”
+712, “Calibration failed”
+720, “Auto-calibration failed; input signal detected”
+740, “Calibration data lost: secure state”