Specifications

Table Of Contents
Chapter 2 Service
Error Messages
164 Assembly-Level Service Guide
Hardware Errors (-240, -241, -250...)
-240, “Hardware error; CPU board initialization failed”
–(1)
P0500-63001 processor board
-240, “Hardware error; no communicate with measurement hardware”
(1) 532xx-66501 motherboard; (2)
P0500-63001 processor board
-240, “Hardware error; failed to program measurement FGCA security
EEPROM”
(1) incorrect firmware match for this model number. Contact
factory support to see if EEPROM U600 needs replacing (counter
must be returned to factory).
-240, “Hardware error; measurement hardware initialization failed”
(1) 532xx-66501 motherboard; (2)
P0500-63001 processor board
-240, “Hardware error; cannot communicate with channel 3 hardware”
(1) 53200-66505 (Option 106) or 53200-66506 (Option 115);
(2) 532xx-66501 motherboard
-240, “Hardware error; channel 3 operation failed”
(1) 53200-66505 (Option 106) or 53200-66506 (Option 115);
(2) 532xx-66501 motherboard
-240, “Hardware error; measurement operation failed”
(1) 532xx-66501 motherboard; (2) 53200-66505 (Option 106) or
53200-66506 (Option 115)
-240, “Hardware error; GPIB interface failed”
(1) Remove the cover and check that the GPIB ribbon cable is fully
seated; (2) P0001-20001 GPIB board; (3)
P0500-63001 processor
board; (4) 532xx-66501 motherboard
-241, “Hardware missing”
(1) GPIB, Internal Battery, or Channel 2/3 not installed
-250, “Mass storage error: file read/write error”
(1) If USB, replace USB drive; (2) P0500-63001 Processor board