Error Messages
Some error messages may be caused by not having the latest firmware for
the counter. Before assuming there is a hardware failure, check the
firmware version in your counter and then contact Agilent using the
procedure at the front of this service guide to determine if you need to
update the firmware to a later version. Refer to
“To Determine the Counter
Firmware Version” on page 160.
If the counter has the latest firmware, and the error message(s) repeat,
follow the recommended assembly replacement procedure.
The following error messages may appear during operation, calibration or
testing of the 53210A, 53220A, or 53230A Counters. Refer to the
appropriate category of error messages for problems you are experiencing
with your counter.
NOTE When information follows an error message, (1) is the most likely
cause of the failure, and then (2), (3), etc. are the next most likely
causes (in that order). Replace the indicated failed assembly. See
Chapter 3 “Replacing Assemblies” and Chapter 5 “Replaceable
Parts”.
Command Errors (-100...)
These errors indicate operator errors. Refer to the User’s Guide for
information.
Execution Errors (-200...)
These errors (except for the following) indicate remote program command
errors. Refer to the SCPI Programmer’s Guide for information.