Specifications

Table Of Contents
Chapter 1 Performance Tests
Introduction
16 Assembly-Level Service Guide
Equipment Required
Table 1-1. Recommended Test Equipment For Operational Verification
and Performance Tests
Instrument Required
Characteristics
Recommended Model
Microwave Signal
Generator
100 MHz to 15 GHz; Freq
Accy at 6 GHz and 15
GHz:
±10ppb; PW Accy:
±10%
Agilent E8257D-520-UNW
or equivalent
Pulse Generator 5 nSec pulse width
±
40 pSec
Agilent 81134A or
equivalent
RF Signal Generator 10 MHz
± 2.000E-12 Stanford Research
Systems CG635
Primary Frequency
Standard (10 MHz House
Standard)
Absolute accuracy >
1 x 10
-10
Stanford Research
Systems FS725 Rubidium
Timebase or equivalent
DC Power Supply Adjustable to 5.000V Agilent 6234A or
equivalent
Digital Multimeter
(DMM)
Microvolt accuracy Agilent 34410A or
equivalent
SMA(m)-to-BNC(f)
Adapters (3)
Agilent 1250-1200
Type N(m)-to-BNC(f)
Adapter
Agilent 1250-0780
BNC-to-Dual Banana Plug BNC(f) to banana plug (m) Agilent E9637A or
Newark 10M4161
50 Coaxial Cable with
BNC connectors (5)
BNC(m) to BNC(m), 48
inches
Agilent 10503A
3.5mm cables (2) DC-18 GHz,matched pair Agilent 11500E
Coaxial BNC Tee
Connector
1250-0781