Specifications

Table Of Contents
Chapter 2 Service
Troubleshooting the Counter
Assembly-Level Service Guide 159
Performing The Counter Self-Test
1 To access the counter Self-Test, press:
[Utility] / (Instr Setup) / (Self Test)
2 When the Self Test softkey is pressed:
The display shows: Performing Self-Test. Please wait.
After several seconds, the display shows: Self-Test Passed.
NOTE If the self-test fails, an error message will be displayed (“Self-Test
Failed.” To check for specific error messages, refer to the
Summary of Remote Self-Test Verification” below for
information on how to read the error message(s) from a remote
computer.
NOTE If you do not have a remote computer available (requires a LAN,
USB or GPIB interface) or are not proficient at programming,
contact the nearest Agilent Service Center for information and
help on sending the counter to Agilent for repair and/or
calibration.