3 Check the battery charge level by pressing the Battery Level softkey.
Display shows (for example):
A battery icon and/or an AC power plug can be seen at the top right-
hand corner of the display, showing power connections and battery
charge level:
NOTE The battery should only be disabled (Off) when the counter is
being stored and unused for an extended period of time. This will
minimize battery drain.
When the battery is enabled and the counter is not connected to AC power,
the battery is discharged at approximately 30% of full initial capacity per
day. When the battery is disabled and not connected to AC power, the
battery is discharged at approximately 10% of full initial capacity per
month.
Batteries that are left fully discharged for 6-18 months may need to be
replaced.
Remote Battery Commands
The battery charge can be read from a remote computer by sending the
SCPI command: SYSTem:BATTery:LEVel? The charge level in % will
be returned.
The counter can be queried to see if it is presently using AC power or
battery power by sending the SCPI command:
SYSTem:BATTery:STATus? The command returns either AC or BATT.