Specifications

Table Of Contents
Chapter 2 Service
Troubleshooting the Counter
Assembly-Level Service Guide 157
Both pins 5 & 6 should read +15 VDC
± 0.2 VDC.
Pins 1 & 2 should also read +15 VDC
± 0.2 VDC. (AC Power supply output)
11 If the voltages in step 8 above are missing, or if there is no reading
on the battery charge level indicator in step 7, replace the battery
(p/n 1420-0909).
NOTE If the battery has been totally discharged for an extended period
of time, you can try charging it for 24 hours to see if it will take a
charge.
12 If the voltages in step 10 are not +15 VDC, replace the Battery
Charging assembly (p/o 53200-80003).
Normal Battery Operation
When the Option 300 battery is installed, it is normally turned ON when
the counter is using AC power, as follows:
1Press the Utility key and then the Battery Option softkey.
2 Press to select:
•Battery: On (When battery is Off, it is disabled but still available.)