Specifications

Table Of Contents
Chapter 2 Service
Troubleshooting the Counter
Assembly-Level Service Guide 155
Remove the two T20 TORX screws and lock washers
holding the battery assembly to the left-hand side panel (as
viewed from the rear of the instrument).
Lift up carefully on the two side-overlapping tabs on the
left side of the battery assembly and separate the assembly
from the three large tabs holding it on the right side. Lay
the assembly on its back to the right side of the chassis and
place the three small tabs (on the back side of the three
large tabs) into the three slots on the assembly deck. (See
Figure 2-12.)
Figure 2-12. Battery folded in the service position
large tabs
small tabs
slots (3)
J100