Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
Assembly-Level Service Guide 141
Resetting the Security Code to a Null
In case a CAL Security Override is required at any time:
Create a text file with the defined name “CalSecureOverride.txt” in an
external USB flash drive at the root directory. This file should be
empty.
Connect the USB drive to the counter front panel USB port.
NOTE Where USB usage is not allowed, FTP can be used to input the file
into the instrument.
Power the instrument off and back on.
At startup, the counter will detect the “CalSecureOverride.txt”
empty file on the USB flash drive and there will be a message on
the front panel display: “CAL Security Override file found. Turn
OFF CAL security?” which must be responded to within 10
seconds time.
If the operator responds to the message (by pressing the YES soft key),
the counter will turn OFF the CALibration SECurity STATe.
A front panel message “CAL Secure State is turned Off” will be
displayed.
As the Calibration secure state is now OFF (i.e.the counter has been
un-secured) the operator can enter a new Secure Code.
At this point, the operator can either turn the Calibration secure state
back ON or leave the box unsecured, as desired.
The “CalSecureOverride.txt” file is automatically erased from the
USB flash drive, regardless of the user action or non-action.