Specifications

Table Of Contents
Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
138 Assembly-Level Service Guide
2Press Auto Cal.
Display stops updating until auto-calibration is completed (several
seconds).
If the calibration is successful, the screen displays “Auto Calibration
succeeded”.
Cal Count is not updated, but the calibration constants ARE
updated.
If any one of the calibration routines fails, the screen displays “Auto
Calibration failed”.
NOTE When the auto-calibration completes, a pass/fail message will be
displayed. If the calibration failed, then the current calibration
constants remain unchanged.
3 If the auto-calibration fails (“Auto-calibration failed”), first verify
any inputs connected to the counter are turned OFF. Then, refer
to the troubleshooting procedures starting on page 150 .
4 To check for error messages, refer to the “
Summary of Remote
Auto-Calibration
” below for information on how to read the
error message(s) from a remote computer.
NOTE If you do not have a remote computer available (requires a LAN,
USB or GPIB interface) or are not proficient at programming,
contact the nearest Agilent Service Center for information and
help on sending the counter to Agilent for repair and/or
calibration.
5 If you are able to read the error message(s), consult “Error
Messages” starting on page 163 for what to do after receiving a
specific calibration error message.