Chapter 1 Performance Tests
Introduction
Assembly-Level Service Guide 13
Chapter Overview
Power-On Test
When the counter is first turned ON, a limited turn-on test is performed to
verify the counter is sufficiently functional to perform basic operations.
Self-Test
Running the counter Self-Test initiates a series of checks against internal
standards. Successful completion of the Self-Test provides a high
confidence level that the counter is completely functional. No external test
equipment is required for this operator-inititated test.
Operational Verification
The operational verification is an abbreviated check that may be
performed to give a strong degree of confidence that the instrument is
capable of making basic measurements accurately without running the
complete performance tests. An operational verification is useful for
incoming inspection and routine preventative maintenance.
Complete Performance Verification Tests
The complete performance verification tests verify the specifications listed
on the Agilent 53210A/53220A/53230A Product Reference CD-ROM as
“Agilent 53210A/53220A/53230A Specifications.” All tests can be
performed without accessing the inside of the instrument. The Complete
Performance Tests are recommended as acceptance tests upon initial
receipt of your counter. They will also help you to become familiar with its
performance and operation. Complete Performance Tests should be run at
least once a year to ensure the counter is meeting all specifications.
Quick Performance Check A combination of selected performance tests
(marked ‘Q’) plus the counter Self-Test provides a simple method to
achieve a high confidence in the counter’s ability to both functionally
operate and to meet specifications. These tests represent the absolute
minimum set of performance checks recommended following any service
activity.