Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
Assembly-Level Service Guide 129
Option 115 Microwave Channel Frequency Calibration
The following procedure should be performed when Option 115 (15 GHz
Microwave channel) with any Option 115 firmware revision is installed as
Channel 2 in a 53210A or Channel 3 in a 53220A or 53230A.
1 Press the following keys:
[Help] / (About)
2 The display shows:
3 Run the counter Self-Test as described under “Performing The
Counter Self-Test” on page 159. A successful self-test procedure
verifies that the counter is operational and ready for calibration.
4 Unlock the counter by performing the procedure “To Unlock The
Counter For Calibration:” on page 103. Key sequence:
[Utility] / (Cal AutoCal) / enter <code> / (Unlock Cal)
5 Set up an external microwave frequency source for a 300 MHz
sine wave at -21 dBm. If the source has an ‘output enable’ feature,
turn it OFF at first (for the zero-calibration procedure).