Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
Assembly-Level Service Guide 127
20 To check for error messages, refer to the “
Summary of Option
106 Remote Microwave Calibration
” below for information on
how to read the error message(s) from a remote computer.
NOTE If you do not have a remote computer available (requires a LAN,
USB or GPIB interface) or are not proficient at programming,
contact the nearest Agilent Service Center for information and
help on sending the counter to Agilent for repair and/or
calibration.
21 If you are able to read the error message(s), consult “Error
Messages” starting on page 163 for what to do after receiving a
specific calibration error message.
Summary of Option 106 Microwave Calibration Procedure:
Input signal: This should be a clean sine wave at each frequency and
power level, driving 50 termination.
Timebase: Any external timebase you provide is ignored during
calibration.
Verify signal input connections and source output is ON.
Procedure: From the front-panel, the following key strokes initiate
the microwave frequency channel calibration:
[Utility] / (Cal AutoCal) / (Perform Cal) / (Cal Step uWave) / enter (Input
Freq) <frequency>
[Utility] / (Cal AutoCal) / (Perform Cal) / (Cal Step uWave) / enter (Input
Power) <power> / (Perform Cal Step)
Summary of Remote Option 106 Microwave Calibration :
Set timeout for PC to be greater than 20 seconds.
Unlock security: CALibration:SECurity:STATe OFF, <code>
Read security state: CALibration:SECurity:STATe?
0 means OFF or unlocked.