Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
Assembly-Level Service Guide 125
After a few seconds, display momentarily shows:
13 Select the ‘output enable’ ON feature of the microwave source or
reconnect the cable to the counter, as appropriate.
14 Use the front panel knob and left/right arrow keys to enter the
value (500 MHz) as shown.
15 Press the Input Power softkey and enter the power level of the 500
MHz source as -27 dBm (actual input power at the counter must
be -27 dBm ± 0.5 dB).
16 Press the Perform Cal Step softkey to enter the data into the
counter and initiate the microwave oscillator calibration.
If the calibration step PASSES, after a few seconds the counter
momentarily displays
“Calibration step succeeded.”
The internal calibration constants are updated.
–The Cal Count on the screen will increment.