Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
Assembly-Level Service Guide 115
Voltage Calibration (Internal DACs)
NOTE The Calibration menu is accessed by pressing the Utility key and
then the Cal/AutoCal softkey.
1 First, run the counter Self-Test as described under “Performing
The Counter Self-Test” on page 159. A successful self-test
procedure verifies that the counter is operational and ready for
calibration.
2 Unlock the counter by performing the procedure ““To Unlock The
Counter For Calibration:” on page 103. Key sequence:
[Utility] / (Cal AutoCal) / enter <code> / (Unlock Cal )
3 Set up an external, low-noise +/- DC voltage source, such as the
Agilent 6234A DC Power Supply, to output a voltage of +4.0 VDC.
Do not connect the output to the counter input connector until
instructed to do so.
NOTE The preferred voltage for optimum calibration accuracy is +4.096
Volts DC. However, if your power supply can not be set to that
specific value, set it as close to +4.0 VDC as possible (acceptable
range is 3-5 V). The entered value for calibration MUST be within
1.0 milliVolt of the actual voltage source output. It is suggested
that a BNC ‘T’ be connected to the power supply output to enable
a DMM to measure the voltage that will be applied to the counter.
4Press the Utility key and then the Cal/AutoCal softkey: