Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
Assembly-Level Service Guide 111
7 If “Unlock Cal” is displayed on a softkey, the counter is locked
against calibration. (See step 2.)
8 If “Lock Cal” is displayed on a softkey, the counter is unlocked
and ready for calibration.
•The Perform Cal softkey becomes available.
9Press the Perform Cal softkey:
The Cal Step softkey shows the last cal step that was selected:
10 Press the Cal Step softkey.
Display softkey annunciators show:
* Only displayed if Option 106 or Option 115 is installed.
11 Press the Oscil softkey.
Display softkey annunciators show:
NOTE The data entry area will display the expected calibration
frequency (10.000 MHz).
*