Specifications

Table Of Contents
Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
Assembly-Level Service Guide 105
The Agilent 53210A/53220A/53230A
Calibration Procedures
Equipment Required
Table 2-1. Recommended Test Equipment For Calibration
Instrument Required Characteristics Recommended Model
10 MHz System Clock or
RF Signal Generator
10 MHz ± 2.000E-12 Stanford Research
Systems CG635 or
equivalent
Primary Frequency Standard
(10 MHz House Standard)
Absolute accuracy >1 x 10
-10
Stanford Research
Systems FS725 Rubidium
Timebase or equivalent
Digital Multimeter (DMM) Microvolt accuracy Agilent 34410A or
equivalent
Microwave Signal Generator 100 MHz to 15 GHz; Freq Accy
at 6 GHz and 15 GHz:
±10ppb; PW Accy: ±10%
Agilent E8257D-520-UNW
or equivalent
DC Power Supply Adjustable to ± 5.000V Agilent 6234A or
equivalent
Type N(m)-to-BNC(f) Adapter Agilent 1250-0780
SMA(m)-to-BNC(f) Adapter Agilent 1250-1200
BNC-to-Dual Banana Plug BNC(f) to banana plug(m) Agilent E9637A or Newark
10M4161
50 Ohm Coaxial Cable with
BNC connectors (5)
BNC(m) to BNC(m), 48 inches Agilent 10503A
3.5mm cables DC-18 GHz, matched pair Agilent 11500E
Coaxial BNC Tee Connector 1250-0781
Some of this equipment may no longer be available as new.
Use an equivalent model number.