Chapter 2 Service
About the Agilent 53210A/53220A/53230A Calibration Menu
Assembly-Level Service Guide 103
To Unlock The Counter For Calibration:
1Press the Utility key, and then press the Cal/AutoCal softkey.
NOTE If Lock Cal is displayed, then the counter is already unlocked for
calibration, and the remaining steps are not applicable.
Unlock For Calibration:
1If Unlock Cal is displayed, use the counter front panel knob and the
appropriate left/right arrow keys to enter a valid security code into
the data entry field on the display (see Figures 2-1 and 2-2).
For example, if AT53210A was the security code set at the factory (for
a 53210A Counter), then enter AT53210A into the data entry field.
The code will repeat in the Secure Code line in the Calibrate
Information screen.