Chapter 2 Service
About the Agilent 53210A/53220A/53230A Calibration Menu
Assembly-Level Service Guide 101
STRing. Refer to the Agilent 53210A/53220A/53230A SCPI
Programmer’s Reference for details of the command format.
Secure Code displays either all dashes ( --- ) or an operator-
entered security code that has not yet been accepted by the
counter.
Display softkey annunciators show:
If “Unlock Cal” is displayed for the second softkey from the
left, the counter is locked against calibration. (See step 1.)
If “Lock Cal” is displayed for the second softkey from the left,
the counter is unlocked and ready for calibration.
To Lock The Counter Against Calibration
NOTE When the counter security is locked to prevent unauthorized
calibration, the Unlock Cal softkey is available. When the counter
security is unlocked, the Lock Cal softkey replaces the Unlock
softkey.
Display softkey annunciators show: