Instruction manual

Chapter 4 Command Reference
Introduction
4-2 Programming Guide
4
Introduction
This chapter describes the SCPI Subsystem commands and the
IEEE 488.2 Common Commands for the Agilent 53150A/151A/152A.
The information in this chapter is intended to help you program the
Counter over its GPIB or RS-232 serial interface.
The commands are presented in alphabetical order.
SCPI Subsystem commands are described on pages 4-4 through 4-42.
IEEE 488.2 Common command descriptions start on Page 4-43.
A description of the Group Execute Trigger command is also included
on Page 4-7.
For each command description:
Where the phrase “Sets or queries” is used, the command setting can
be queried by omitting the parameter and appending a “?” to the last
command keyword.
For example,
:ROSC:SOUR INT | EXT
can be queried with
:ROSC:SOUR?
Unless otherwise noted, a command described as an event cannot
be queried.
Unless otherwise noted, the command setting is affected by *SAV/*RCL.
The square brackets, [ ], are used to indicate that the element(s) within
the brackets are optional. Note, the brackets are NOT part of the
command and should not be sent to the Counter.