Instruction manual

Chapter 3 Programming Your Counter for Remote Operation
Status Reporting
Programming Guide 3-37
3
Operation Status Register Group
The Operation Status Register Group monitors conditions which are part
of the Counter’s normal operation and has a complete set of registers that
consist of the following:
a condition register
a positive transition filter register (PTR)
a negative transition filter register (NTR)
an event register
an event enable register
Table 3-7 lists the Operation Status Register bits and briefly describes
each bit.
Figure 3-13 shows the model that these register groups follow.
Table 3-7. Operation Status Register
BIT WEIGHT DESCRIPTION
0
Not used
1 - 3
Not used
4
16 Measuring
5
32 Waiting for Trigger
6 - 8
Not used
9
512 Using Internal Reference
10
Not used
11
Acquiring
12
Locked
13 - 14
Not used
15
Not used, since some controllers may have difficulty reading
a 16-bit unsigned integer. The value of this bit is always 0.