Transition filters are unaffected by *CLS or queries. Transition filters are
set to default values by :STATus:PRESet and power-on.
To write to the Operation Status transition filter registers use:
:STATus:OPERation:PTRansition
:STATus:OPERation:NTRansition
To read these registers use:
:STATus:OPERation:PTRansition?
:STATus:OPERation:NTRansition?
Event Register
An event register captures changes in conditions.
An event register bit (event bit) is set TRUE when an associated event
occurs. These bits, once set, are “sticky.” That is, they cannot be cleared
even if they do not reflect the current status of a related condition, until
they are read.
Table 3-6. Transition Filter Definition
Positive
Transition
Filter Bit
Negative
Transition
Filter Bit
Transition Which Causes the Event-Bit to be set TRUE
TRUE FALSE positive transition
FALSE TRUE negative transition
TRUE TRUE either a positive or negative transition
FALSE FALSE neither transition (event reporting is disabled)