Instruction manual

Chapter 3 Programming Your Counter for Remote Operation
Status Reporting
Programming Guide 3-31
3
A detailed description of each bit in the Standard Event Status Register
follows:
Bit 0 (Operation Complete) is an event bit which is generated in
response to the *OPC command. This bit indicates that the Counter
has completed all pending operations (the pending operation condition
has transitioned from TRUE to FALSE).
If AVERage:STATe is OFF, the command INIT;*OPC sets the OPC bit
once the instrument completes a measurement; if AVERage:STATe is
ON, the command INIT;*OPC sets the OPC bit once the instrument
completes a measurement consisting of AVERage:COUNt measurements.
NOTE The OPC bit is not in any way affected by the *OPC? query.
Bit 1 is not used.
Bit 2 (Query Error) is an event bit which indicates that either 1) an
attempt was made to read the Output Queue when it was empty or 2)
data in the Output Queue has been lost.
Errors -400 through -499 are query errors.
Table 3-5. Standard Event Status Register
BIT WEIGHT SYMBOL DESCRIPTION
0
1 OPC Operation Complete
1
(RQC) Not used because this instrument cannot request permission
to become active IEEE 488.1 controller-in-charge.
2
4 QYE Query Error
3
8 DDE Device-Specific Error
4
16 EXE Execution Error
5
32 CME Command Error
6
(URQ) Not used, because this instrument does not define any local
controls as “User Request” controls.
7
128 PON Power On