Instruction manual
Chapter 4 Command Reference
:TRIGger Subsystem
4-42 Programming Guide
4
:TRIGger Subsystem
COMMAND :TRIGger[SEQuence | STARt]:HOLDoff . . . <numeric_value>
When INIT:CONT is ON, this command determines the rate at which
measurements are made by setting a delay between measurements.
Its setting corresponds to the front-panel rate setting as follows:
<NUMERIC_VALUE>
RATE
0.0 = FAST; 0.5 = MEDium; 1.0 = SLOW
COMMENTS • *RST: 0.0
• The only settings that are accepted are the three listed above.
• The measurement rate is also affected by the Resolution setting.
• When INIT:CONT is OFF, the holdoff has no effect.
• There is no setting for this command that corresponds to the front-
panel rate setting of HOLD. The HOLD mode is entered through *RST
or INIT:CONT OFF. The instrument then stops making measurements
until it is triggered or until it receives INIT:CONT ON.
RELATED
FRONT-PANEL
KEYS
Rate