Instruction manual
Chapter 4 Command Reference
:STATus Subsystem
Programming Guide 4-35
4
COMMENTS • The parameter and the query response value, when rounded to an
integer value and expressed in base 2 (binary), represent the bit values
of the positive transition filter.
• The value of unused bits is zero when queried and is ignored when set.
• A TRUE bit in the positive transition filter specifies that a positive
(FALSE to TRUE) transition of the corresponding bit in the Operation
Condition Status Register generates the corresponding event in the
Operation Event Status Register.
• At power-on and STAT:PRES, the positive transition filter is preset
such that each bit is a 1 (TRUE).
• This value is unaffected by *RST, *CLS, and *SAV/*RCL.
RELATED
FRONT-PANEL
KEYS
None
COMMAND :STATus:PRESet
This event command presets the enable registers and transition filters
associated with the Operation and Questionable status reporting
structures. The enable registers and negative transition filters are preset
such that each bit is a 0 (FALSE). The positive transition filters are preset
such that each bit is a 1 (TRUE).