Chapter 4 Command Reference
:STATus Subsystem
4-32 Programming Guide
4
COMMAND :STATus:OPERation:CONDition?
Queries the status of the Operation Condition Status Register.
QUERY RESPONSE Numeric data transferred as ASCII bytes in <NR1> format.
Range is 0 to 32,767.
The query response value is an integer formed by the binary weighting
of the bits. The value of unused bits is zero.
COMMENTS Query only.
The Operation Condition Status Register is cleared at power-on.
Bits are not cleared when read.
RELATED
FRONT-PANEL
KEYS
None
COMMAND :STATus:OPERation:ENABle <non-decimal numeric> | <NRf>
Sets or queries the Operation Event Status Enable Register.
RANGE The range for the <non-decimal numeric> or <NRf> parameter is
0 to 32,767.
QUERY RESPONSE Numeric data transferred as ASCII bytes in <NR1> format.
COMMENTS The parameter and query response value, when rounded to an integer
value and expressed in base 2 (binary), represent the bit values of the
Operation Event Status Enable Register.
The value of unused bits is zero when queried and is ignored when set.
This register is used to enable a single or inclusive OR group of
Operation Event Status Register events to be summarized in the
Status Byte Register (bit 7).
At power-on and :STAT:PRES, the Operation Event Status Enable
Register is cleared (value is 0).
This value is unaffected by *RST and *SAV/*RCL.
RELATED
FRONT-PANEL
KEYS
None